RP-Series

RP-4500D2

Differential Input Panel Meter

4 1/2 Digit 0.56 LED

in a NEMA Style Case

RP-4500D2BCD

Meter with Parallel BCD Output

A High Accuracy Differential Input

AC Powered Meter, and a Tri-State Buffered Parallel BCD Output Option.

General Features

The Texmate Model RP-4500D2 is an ultra stable, super accurate

4 1/2 digit panel meter featuring a 100KHz crystal clock. As a result of this feature, the meter provides an exceptionally high normal mode and common mode rejection. The meter measures bipolar true differential and single-ended DC voltages over four factory calibrated ranges from 1.9999V to 1200.0V full scale.

Resolution is 100V over 19999 counts, and errors due to zero drift are virtually eliminated by means of autozeroing. The meter has a multiplexed BCD output capability and as an option a parallel BCD output can be provided. Other user programmable modes of operation include an ohmmeter, current meter, ratiometric voltmeter, and special scaling with offset capability.

The differential input capability of the RP-4500D2 has a wide common mode voltage range of 3VDC and a high common mode DC voltage rejection ratio of 80dB. Common mode signals are those which are present equally on both input terminals but do not develop a differential voltage between them. This capability is particularly useful for making accurate measurements of very small signals in the presence of much larger common mode signals. Because of its noise immunity, the RP-4500D2 is ideal for measuring various balanced transducers and bridge inputs.

The 120dB normal mode rejection of the RP-4500D2 at multiples of 50/60Hz means that almost any AC mains noise present on the input signal will be rejected.

In addition to standard run/hold and display blanking options, the RP-4500D2 features specially prepared internal mounting holes and solder pads to enable user designed input signal conditioning and control. These auxiliary pads provide access to all the key operating and control circuitry of the meter including regulated 5VDC and outputs.

Specifications and features of the RP-4500D2BCD meter begin on page 5.

Compatibility

The RP-Series NEMA case style is complementary to Texmates Classic UM-Series.

For economy, each RP model is dedicated to a specific application. RPs are ideal for upgrading or replacing the traditional USA NEMA case panel meters presently in use.

Traditional

NEMA

STYLE USA

CASE

Specifications

Input Configuration:......True differential and single-ended Full Scale Ranges:........199.9mVDC

1.999VDC (standard)

19.99VDC

199.9VDC

1200VDC

Input Impedance:..........Exceeds 1000M on 2V range;

10M on all other ranges

Input Protection:............60VDC or 40VAC on 2V range;

1200VDC or 850VAC on all other ranges

Accuracy:......................(0.01% of reading + 3 digits) Temperature Coefficient:..5PPM/C in ratiometric operation; 50 PPM/C Typ. using internal reference with 2V range

Warm Up Time:..............3 minutes to specified accuracy Conversion Rate:..........2.5 readings per second controlled

by precision 100KHz Quartz Crystal Clock, or user controllable from 1 to

12 readings per second by external clock

Display:..........................0.56" LED

Decimal Selection:........User programmable to 4 positions Overrange Indication:..When input exceeds full scale on

any range being used, display flashes 000

Power Requirements:....110V or 220V, 5% at 50Hz; 117V

or 230V, 5% at 60 and 400Hz

Operating Temperature:....-10 to +50C Storage Temperature:....-20 to +70C Relative Humidity..........95% (non-condensing) Case Dimensions:........Bezel 4.06Wx1.89H (102.7Wx47.9Hmm) Depth behind bezel 3.64" (92.22 mm) Plus 0.5 to.9 (12.7 to 22.8mm) depending on connector used.

Weight:............................8 oz (227 gms)

RP-Series, a reliable replacement for your application

RP-3500D2................3.5 digit Red LED Ultra Stable, Differential, 2VDC std RP-3500D2BCD........RP-3500D2 with Tri-State Parallel BCD, 2VDC std RP-35A......................3.5 digit Red LED with Differential Inputs, 2VDC std RP-35AR....................3.5 digit Red LED, Autoranging, 200mV / 2VDC

RP-35U......................3.5 digit Red LED, Low Cost, 2VDC std RP-4500D2................4.5 digit RED LED Ultra Stable, Differential, 2VDC std RP-4500D2BCD........RP-4500D2 with Tri-State Parallel BCD, 2VDC std

Functional Diagram

Connector Pinouts

The Texmate Model RP-4500D2 interconnects by means of a standard PC board edge connector having two rows of 15 pins each, spaced on 0.156" centers. Connectors are available from Texmate, or from almost any connector manufacturer.

Rear View of Meter

PCB

PINS A, B, 1 and 2 - Decimal Select: Decimal points may be displayed as required by connecting appropriate pin to System Ground Pin K.

PIN C - Constant Voltage Output: Pin C may be used as a primary reference Voltage Source of +2.5V referred to Analog Common Pin D. A temperature coefficient of 50PPM/ C is typical. The maximum usable load without component change is limited to 500A.

PIN D - Analog Common: Signal return common for differential inputs, ratiometric inputs, external reference inputs. For single-ended inputs, Pin D and Pin 4 should be joined at the grounded side of the input signal source. CAUTION: Analog Common Pin D is not isolated and is internally connected to System Ground Pin K.

PIN E - Signal High Input: Signal high input of A to D converter. Maximum overvoltage protection is 60VDC or 40VAC.

PIN F - Reference Voltage Output: Internal precision voltage reference. Standard output is 1.000V, adjustable by 5% with R10 potentiometer. Usable voltages from 0.5V to 2.5V for special high impedance scaling can be obtained by changing the value of internal dividing resistors R8 and R9.

PIN H - Display Power Input: Power input for LED display

drive. For normal operation, connect Pin H to +5VDC Power Output Pin 7. The display may be dimmed or blanked by reducing or removing the voltage between Pin H and Pin 7. The power supply to the A to D converter and logic circuits is independent to that of the display. Thus, even with the display blanked, the remainder of the meter continues to function normally.

PIN J - Clock Output: A quartz crystal controlled oscillator provides a stable clock signal output of 100KHz.

PIN K - System Ground: All digital signals, decimal points, display test, and run/hold circuits should be returned to this ground point. Pin K is internally connected to Analog Common Pin D.

PIN L - Run/Hold: If Pin L is left open (or connected to +5VDC Power Output Pin 7 for logic control purposes), the meter will operate in a free-running mode. Under control of the internal 100KHz quartz crystal clock, readings will be updated every 400mS (2.5 per sec). If Pin L is connected to System Ground Pin K (logic low), the meter will latch up an continuously display the reading. If Pin L is released from Pin K (Pin L then goes logic high) for more than 2000 clock pulses (>10mS at 100KHz) and returned to Pin K (logic low), the meter will complete one conversion, update, and then hold the new reading. For all practical purposes, a manually actuated pushbutton switch will provide sufficient timing for press-to-update operation.

PIN M - Integration Period BUSY Output: CMOS logic output (normally high) goes low and remains low during the period of the conversion cycle when the input signal is being integrated.

Actual integration is performed for 20,000 clock pulses. De-integration then occurs for 0 to 40,000 clock pulses (each count display = 2 clock pulses), depending on signal magnitude. Pin M goes to logic high throughout Autozero Phase and during Hold Status.

PINS N, R, 11, 13, and 15 - No Connection: The PCB pads which would normally correspond to these pins do not exist on the PCB.

Pin P - 117/230V Select: Connect Pin P to AC Power Input Pin 12 for 117V operation. Connect Pin P to 117/230V Select Pin 14 for 230V operation.

PIN S - AC Power Input: Connect one side of 117 or 230VAC power input to Pin S.

PIN 3 - Offset Voltage Output: Pin 3 is the center tap of R15 potentiometer which when installed will provide an offset voltage output which is variable from 0 to +2.5V and may be used for signal offset or scaling purposes.

PIN 4 - Signal Low Input: Signal low input of A to D converter.

Maximum overvoltage protection is 60VDC or 40VAC.

PIN 5 - Voltage Divider Signal High Input: Signal high input for voltages that require attenuation or scaling. Dividing resistors R1 and R2 may be mounted internally for voltages up to 1200V max. Matched dividing resistors for 20V (1/10), 200V (1/100) and 1200V (1/1000) ranges are available from Texmate.

Shunt resistors for current measurements up to 200mA may be internally mounted in the R1 position. The current loop is then applied to Signal High Input Pin E and returned through Pin 5.

PIN 6 - Reference Input: Reference voltage input for A to D converter. Normally supplied from Reference Voltage Output Pin F. An external reference source referred to Analog Common Pin D may be used instead. Pin 6 may be used as an input for ratiometric measurements. Minimum usable voltage is 0.5VDC and the maximum voltage is 2.5VDC. (Signal Input Voltage Reference Input Voltage) x 1000 = Reading Displayed.

Maximum signal voltage is 2.5V. Higher voltages must be scaled down through voltage divider. Reference input voltage must remain stable during measurement period.

PIN 7 - +5VDC Power Output: Regulated +5VDC 3% power output to enable the display and logic circuitry. In addition, up to 25mA may be used to power external circuitry.

PIN 8 - Clock Input: Normally Pin 8 is connected to Clock Output Pin J which provides a 100KHz clock input for optimum rejection from 50/60Hz noise. An external clock source may be used if desired (+5V referred to power ground at 50% duty cycle). The minimum recommended frequency is 10KHz,and the maximum is 1MHz (12.5 readings per sec.). For inputs below 100KHz or above 300KHz, the integrator time constant and some component values have to be changed.

PIN 9Polarity Enable: Connect Pin 9 to System Ground Pin K to enable normal indication of positive and negative polarities. To reverse the polarity indication (i.e. indicate negative for positive inputs and vice versa) open jumper J4 and close jumpers J5 and J8 on the solder side of main board.

PIN 10 - Display Test: All display segments will operate when Pin 10 is connected to System Ground Pin K.

PIN 12 - AC Power Input: Connect one side of 17 or 230 V AC power input to Pin 12.

PIN 14 - 117/230V Select: Connect Pin 14 to AC Power Input Pin S for 117V operation. Connect Pin 14 to 117/230V Select Pin P for 230V operation.

Component Layout

COMPONENT SIDE

SOLDER SIDE

Signal Conditioning Components

SPAN Potentiometer (Pot)

The SPAN pot is on the right side of the display. Typical adjustment is 20% of the input signal range.

ZERO Potentiometer (Pot) optional

The ZERO pot is on the right side of the SPAN Pot. Typically it enables the displayed reading to be offset 500 counts.

Timing Diagram

Calibration Procedure

Apply power to the meter. Then with a precision DC reference source, apply +1.9000VDC between the Signal High Input Pin E and the Signal Low Input Pin 4. Adjust R10 potentiometer (behind front panel filter on right side as viewed from front) until the display reads +1.9000V. Note: The voltage applied in this case is for a +1.9999V full scale meter. For other ranges, the voltage applied should be similarly proportionate to the particular full scale voltage.

Typical Application Connections

The RP-4500D2 may be used in a wide variety of configurations. The following circuits illustrate some of the possibilities and demonstrate the exceptional versatility of Texmate products. Components called for in the applications which are not part of the standard meter may be supplied by the user or in some cases purchased from Texmate. The circuit diagrams explain the basic pinout connections required for each application. Unless otherwise specified, the diagrams will show the component values and solder junctions that would normally be installed on a standard 2V range meter. For those applications which have alternative ranges and/or input configurations, the required component values and any modifications are described in the text.

DIFFERENTIAL METER: 2V, >2V RANGE OR EXT. REF.

2V Range:

1) Connect up the circuit as shown.

2) Apply Signal High to Pin E and Signal Low to Pin 4.

Continued on page 4.

Typical Application Connections continued

SINGLE ENDED CURRENT METER 1) Connect meter as shown. 2) Install RS.

Note: RS must be externally mounted when current is greater than 200mA. Standard values of RS are specified under section titled Useful Tables. For all other ranges, Rs may be internally mounted in the R1 position.

DIFFERENTIAL RATIOMETRIC OHM METER 1) Remove (open circuit) R8. 2) (Rs + RT) x 2 = Full Scale Value.

3) Reading = Rx (Rs + RT) x 10,000. 4) Change R6 from 220k to 22k for 200. 5) Remove C5 for 20M range.

SINGLE ENDED METER: 2V, >2V RANGE For 2V Range:

1) Connect up the circuit as shown.

2) Apply Signal High to Pin E and Signal Low to Pins 4, D.

For >2V Range:

1) Install R1 and R2 as specified under section titled Useful Tables.

2) Apply Signal High to Pin 5 and Signal Low to Pins 4, D.

RATIOMETRIC VOLTMETER

1) Reading = (V1 + V2) x 10,000, where -2V <V1< +2.8V and +100mV <V2< +5V. If 50mV <V2< 500mV, change R6 from 220k to 22k.

SIMULTANEOUS VOLTAGE AND CURRENT MEASUREMENT 1) Install R1, R2 and Rs as specified under section titled Useful Tables.

Note: RS must be located in the low side of the current loop and Signal Low Input Pin 4 of Voltmeter must not be grounded.

DISPLAY BRIGHTNESS CONTROL

Externally mounted diodes may be used to vary display brightness.

DPM

Useful Tables

VOLTAGE RANGE CHANGE

F.S. In

20V

200V

1200V

Omit

9M

10M

10M

Omit

1M

100K

10K

Resol.

100V

1mV

10mV

100mV

CURRENT RANGE CHANGE (*) F.S. In Rs Resol.

0.1 20A 1mA

1 2A 100A

10 200mA 10A

100 20mA 1A

1K 2mA 100nA

10K 200A 10nA

100K 20A 1nA

RESISTANCE RANGE CHANGE

F.S. In

200

2K

20K

200K

2M

20M

Rs + RT

100

1K

10K

100K

1M

10M

Resol.

10m

100m

10

100

1K

* For full scale current inputs greater than 200mA, Rs must be located external to the meter, and a 4-wire type connection should be used. For full scale currents of 200mA or less, Rs may be located on the meter's PCB in the R1 position.

Parallel BCD Output Option

General Features

The RP-4500D2BCD meter, using a parallel BCD Output Module, is designed to further increase the versatility of the RP-4500D2 meter. it consists of an additional PC board mounted in the case with the meter and connected to it with a multiconductor ribbon cable.

The Parallel BCD module incorporates CMOS circuitry wish Tristate output, capable of two low power TTL loads.

The unit may be used to drive remote displays, digital printers, digital comparators, or provide data for digital processor applications.

Specifications and features of the RP-4500D2 meter begin on page 1.

Functional Diagram

Compatibility

The RP-Series NEMA case style is complementary to Texmates Classic UM-Series.

For economy, each RP model is dedicated to a specific application. RPs are ideal for upgrading or replacing the traditional USA NEMA case panel meters presently in use.

Traditional

NEMA

STYLE USA

CASE

Specifications

System Voltage:............5VDC 2% Data Output:..................Parallel binary coded decimal (BCD)

BCD Logic Level:..........Buffered CMOS tri-state output

19999 counts max.

Logic 1: 4.6V at 1mA Logic 0: 0.4V at 1mA Logic X: High impedance 300nA max

Polarity Output:............Logic 1 for positive Logic 0 for negative

Spare Data Latch:..........2 bits spare latch available for deci-

mal points, overrange, and underrange outputs.

Data Input:......................Multiplexed BCD CMOS input Control Output:..............Data ready, print command, overrange, underrange, busy, run/hold, clear, blanking, and output select.

Power Consumption:....2mA typical at no output load Input Configuration:......True differential and single-ended Full Scale Ranges:........199.9mVDC

1.999VDC (standard)

19.99VDC

199.9VDC

1200VDC

Input Impedance:..........Exceeds 1000M on 2V range;

10M on all other ranges

Input Protection:............60VDC or 40VAC on 2V range;

1200VDC or 850VAC on all other ranges

Accuracy:......................(0.01% of reading + 3 digits) Temperature Coefficient:..5PPM/C in ratiometric operation; 50 PPM/C Typ. using internal reference with 2V range

Warm Up Time:..............3 minutes to specified accuracy Conversion Rate:..........2.5 readings per second controlled

by precision 100KHz Quartz Crystal Clock, or user controllable from 1 to

12 readings per second by external clock

Display:..........................0.56" LED

Decimal Selection:........User programmable to 4 positions Overrange Indication:..When input exceeds full scale on

any range being used, display flashes 000

Power Requirements:....110V or 220V, 5% at 50Hz; 117V

or 230V, 5% at 60 and 400Hz

Operating Temperature:....-10 to +50C Storage Temperature:....-20 to +70C Relative Humidity..........95% (non-condensing) Case Dimensions:........Bezel 4.06Wx1.89H (102.7Wx47.9Hmm) Depth behind bezel 3.64" (92.22 mm) Plus 0.5 to.9 (12.7 to 22.8mm) depending on connector used.

Weight:............................8 oz (227 gms)

Connector Pinouts

printer, and then returns to logic 1.

The Texmate Model RP-4500D2BCD interconnects by means of a standard PC board edge connector having two rows of 15 pins each, spaced on 0.156" centers. Connectors are available from Texmate, or from almost any connector manufacturer.

Pins 2 and 3 - Spare Inputs: Any required digital information (2 bits), such as decimal point positions, overrange / underrange can be transmitted to the data bus through Pin B and Pin C by using these two latches.

Rear View of Meter

PCB

Pin 5 - Overrange Output: Pin 5 is normally logic 0 (zero volts). At the end of the deintegration period if the displayed reading is greater than 19999, Pin 5 will go 1 (+V). Pin 5 will change state from 1 to 0 at the beginning of the deintegration period.

Pin A - Data Ready Output: Pin A remains at logic 0 during the period that valid data is available. Pin A goes high (logic 1) for approximately 20mS after the integrator output crosses zero. During this period of change, the output data is considered invalid.

Pins B and C - Spare Outputs: Pins B and C are spare TriState latched outputs that can be addressed by the spare inputs provided on Pins 2 and 3.

Pin D - Polarity Sign Output: Pin D is a latched Tri-State output. Its output data is updated after every conversion cycle. Pin D goes to a Tri-State condition if Output Enable Pin R is either left open or applied to logic 1.

Pin E - System Ground: Pin E is the common for all digital signals.

Pins F, H, J, K, L, M, N, P, 4, 6, 7, 8, 9, 10, 11, 12 and 13 - Data Outputs: These pins are latched Tri-State outputs.

Data is updated after every conversion cycle. They go to a Tri-State condition if Output Enable Pin R is either left open or applied to logic 1.

Pin R - Output Enable: For single meter applications, apply logic 0 (or connect to System Ground Pin E) to enable all data outputs. For multiple meter applications using a single data retriever such as a printer, apply logic 1 to Pin R in order to produce a floating output (Tri-State) at the Data Output Pins F, H, J, K, L, M, N, P, 4, 6, 7, 8, 9, 10, 11, 12 and 13.

Pin S - Run / Hold Input: Apply logic 0 to Pin S to hold the input. Apply logic 1 for free-running operation.

Pin 1 - Print Command Output: Pin 1 is normally at logic 1 and goes to logic 0 for approximately 20mS to activate a

Pin 14 - Clear Input: By applying logic 1 to this pin, all data outputs will be cleared to zero. Application of logic 0 or an open circuit will enable normal operation.

Pin 15 - Busy Operation: Pin 15 is normally logic 1. It goes to logic 0 during the integration period.

BCD Output Board Component Layout

COMPONENT SIDE

SOLDER SIDE

Parallel BCD Output Typical Application Circuits and Connection Instructions The RP-4500D2BCD may be used in a wide variety of configurations. The following circuits illustrate some of the possibilities and demonstrate the exceptional versatility of Texmate products. Components called for in the applications which are not part of the standard meter may be supplied by the user or in some cases purchased from Texmate. The circuit diagrams explain the basic pinout connections required for each application. Unless otherwise specified, the diagrams will show the component values and solder junctions that would normally be installed on a standard 2V range meter. For those applications which have alternative ranges and/or input configurations, the required component values and any modifications are described in the text.

BUS SYSTEM FOR MORE THAN ONE METER USING ONE PRINTER

THUMBWHEEL SWITCH DIGITAL COMPARATOR WITH HIGH, LOW AND EQUAL OUTPUT

Optional PCB Edge Connector PCB Edge Connector

A standard 30-pin edge connector (two rows of 15 pins on 0.156" centers) may be used to connect the RP-Series of meters. Order part no. CN-L15.

Face Plate Descriptors

To customize the face plate, each RP-meter is supplied with a white printed clear adhesive label containing various popular descriptors.

Choose the descriptor, peel off the adhesive backing and align the descriptor in the lower right corner of the standard face plate.

MORE THAN ONE METER COMPARED TO ONE SETPOINT

Custom Face Plates

Texmate Produces Thousands of Custom OEM Face Plates

Have Texmate Design and produce a Custom Face Plate for your next project!

Custom face plates have a nonrecurring artwork charge. A serial number is then assigned to each artwork to facilitate reordering.

Small Run or One-Off custom face plates incur an installation charge, and are generally printed on a special plastic film, which is then laminated to custom faceplate blanks as required.

Large Run (250 pieces min): custom face plates are production silk screened, issued a part number, and held in stock for free installation as required by customer orders.

OEMs may also order Custom Meter Labels, Box Labels, Custom Data Sheets and Instruction Manuals.

RP Case Dimensions and Panel Cutouts

For new installations see Optimum Panel Cutout or Panel Cutout for Front Panel Removal

Ordering Information

Standard Options for this Model Number

(cid:2)ACCESSORIES

Part Number.Description

(cid:2)BASIC MODEL NUMBER

RP-4500D2...4.5 digit Red Ultra Stable, Differential, 2VDC std RP-4500D2BCD..4.5 digit Red Ultra Stable, Tri-State Parallel BCD, 2VDC std

Special Options and Accessories

Part Number.Description

(cid:2)SPECIAL OPTIONS (Specify Inputs & Req. Reading)

VDA-200MVFI...200mVDC Range Change VFA-0020V.....20VDC Range Change for 4.5 digit RP Series VFA-0200V.....200VDC Range Change for 4.5 digit RP Series VFA-1200V.....1200VDC Range Change for 4.5 digit RP Series VS-4.5.........Non-Std Range and Scale - 4.5 Digit RP Meters

CN-L15........PCB Edge Connector, Solder Type, Dual row 15 Pins OP-NSEAL/UM..96x48mm clear lockable front cover - NEMA 4X, splash proof

for RP & UM Series (Factory Installed) RP.CASE.......Replacement Case w/Mount Hardware DN.CAS96X24.Din Case 96 X 24 Short Depth Plus Accessories

For product details visit www.texmate.com